ملخص
Clip slippage is a rare occurrence, and the scissoring or torsional failure of aneurysm clips is rarer still. Titanium clips have been implicated in a few such reported cases. The authors report its occurrence while using a fenestrated cobalt alloy clip for an internal carotid artery aneurysm which was identified by intraoperative angiography and rectified by re-applying a non-fenestrated cobalt alloy clip. The possible mechanism of this complication, and measures that may prevent its occurrence, including meticulous dissection, decompression when possible, proper clip selection, and intraoperative angiogram are described.
اللغة الأصلية | English |
---|---|
الصفحات (من إلى) | 179-182 |
عدد الصفحات | 4 |
دورية | Sultan Qaboos University Medical Journal |
مستوى الصوت | 13 |
رقم الإصدار | 1 |
المعرِّفات الرقمية للأشياء | |
حالة النشر | Published - فبراير 2013 |
ASJC Scopus subject areas
- ???subjectarea.asjc.2700???